Athena-control

Quick RIVERNEW

Simple and Easy displacement measurement! Easy displacement measurement in FA and laboratory (test and research) scenes.

Product Overview

Quick RIVERNEW is an eddy current type sensor used for displacement measurement with excellent environmental resistance and stability.  Easy set-up and measurement make it beneficial for FA and lab (test and research) scenes, as it gives highly accurate measurements of displacement with changes in micron units for thickness, height, and position etc., without contacting a target object.

 
  • Eddy current method: Not affected by water, oil, dust etc.
  • IP Rating IPX7
  • Operating temperature -10 ℃ ~ +70  ℃

Features

With this Quick RIVERNEW, one amplifier is good for different measurement ranges and different cable lengths, unlike the conventional eddy-current type converters which were only compatible with certain cable and sensor.  6 different patterns of system configurations are available according to various applications and intended uses, combining one of the three sensors of 2 mm, 5 mm, 10 mm measurement ranges with either of 3 m or 10 m cable lengths and one amplifier.

Simple

Useful in a simple instrument configuration.
Sensor types and extension cables can also be shared by one amplifier.

  • Range free amplifier. Commonly used for any of the 3 measurement ranges of 2mm, 5mm or 10mm.
  • Sensor can be chosen from three measuring ranges of 2 mm, 5 mm, 10 mm.
  • Extendable length between sensor and amplifier; with the use of 7 m extension cable, overall length can be extended to 10 m.

Easy

Used conveniently according to the purpose. Only 3 points calibration are required.

  • One single amplifier can be used to meaure any of the targets materials such as iron, stainless steel, aluminum, coper etc.
  • One single amplifier can be used with any of the 3 ranges (2 mm, 5 mm, 10 mm) of sensors.
  • One single amplifier can be used for either of (3 m or 10 m) the cable length.

※ “Simple adjustment” is made possible with 3-point calibration using the spacers. For more accurate calibration, you can use a calibration kit with a micrometer.

System Configuration

One amplifier can calibrate to any of three types of sensors × either of 3 m or 10 m the system cable length.

Applications

One amplifier can calibrate to any of three types of sensors × either of 3 m or 10 m the system cable length.

Ordering Information

Set model code ①

System cable length 3 m (Set Contents: Sensor, Amplifier, Spacer set for calibration

Items and quantities in set con tents

No.

Descriptions

Model code

Sensor

Amplifier

Extension cable

Spacer set for calibration

For 0 to 2 mm

For 0 to 5 mm

For 0 to 10 mm

Common use of each range

7 m, common for each range

1

2mm range Set
RXT-CL2000
1unit
1unit
1 / 2 mm 1each

2

5mm range Set
RXT-CL0500
1unit
1unit
2.5 / 5 mm 1each

3

10mm range Set
RXT-CL0010
1unit
1unit
5 / 10 mm 1each

4

Multi range Set
RXT-CL2510
1unit
1unit
1unit
1unit
1 / 2 / 2.5 / 5 / 10 mm 1each

Set model code ②

System cable length 3 m or 10 m (Set Contents: Sensor, Amplifier, Extension cable, Spacer set for calibration

Items and quantities in set con tents

No.

Descriptions

Model code

Sensor

Amplifier

Extension cable

Spacer set for calibration

For 0 to 2 mm

For 0 to 5 mm

For 0 to 10 mm

Common use of each range

7 m, common for each range

1

2mm range Set
RXT-S2000
1unit
1unit
1pc.
1 / 2 mm 1each

2

5mm range Set
RXT-S0500
1unit
1unit
1pc.
2.5 / 5 mm 1each

3

10mm range Set
RXT-S0010
1unit
1unit
1pc.
5 / 10 mm 1each

4

Multi range Set
RXT-S2510
1unit
1unit
1unit
1unit
1pc.
1 / 2 / 2.5 / 5 / 10 mm 1each

Single unit model code

No.

Descriptions

Specifications

Single unit model code

1

Sensor
0 ~ 2 ㎜
RXS-02-M030-03

2

Sensor
0 ~ 5 ㎜
RXS-05-M050-03

3

Sensor
0 ~ 10 ㎜
RXS-10-M050-03

4

Amplifier
Common use of each range
RXC-0
5
Extension cable
7 m, common for each range
RXW-07